Changing the Purge Cassette and Fluid Bag

July 20, 2020

Quick Skills Video for Using the Automated Impella Controller

This quick skills video demonstrates the procedure for changing the purge cassette and fluid bag using the Automated Impella Controller (AIC). See how to interact with the AIC as well as how to perform the steps of changing the purge cassette and the purge fluid bag. The procedure should be completed within 90 seconds and all connections must be wet-to-wet to avoid air entering the system.

This video covers:

  • How to select the Change Cassette & Bag procedure on the AIC
  • What supplies are needed to complete the procedure (D5W dextrose bag with heparin, new purge cassette)
  • How to disconnect the yellow luer from the Impella catheter
  • How to open the purge cassette door and remove the old purge cassette
  • How to replace the purge fluid bag
  • How to insert the new purge cassette
  • How to adjust the purge fluid information (purge fluid volume, dextrose concentration, and heparin concentration)
  • How to remove and discard the Y-connector and connect the yellow luer of the purge tubing to the yellow luer on the Impella catheter

Note: This video does not replace the Instructions for Use manual.

Related Content:

AIC-032-17
AIC-0094

Subscribe or join the conversation by following us on Twitter: @ProtectedPCI


To learn more about the Impella® platform of heart pumps, including important risk and safety information associated with the use of the devices, please visit: abiomed.com/important-safety-information

Subscribe to the Digital Community

Receive weekly updates including case review videos, tips and tricks, specialists opinions, and resources and downloads regarding Protected PCI and Cardiogenic Shock.

Subscribe to the Digital Community

Receive weekly updates including case review videos, tips and tricks, specialists opinions, and resources and downloads regarding Protected PCI and Cardiogenic Shock.